ASM |
|
|
|
Standard Calibration - UV/Visible Spectrophotometers √ Wavelength Accuracy: Certified Holmium Oxide & Didymium Glass Filters √ Wavelength Reproducibility: Certified Holmium Oxide & Didymium Glass Filters √ Photometric Accuracy: Certified Neutral Density Filters √ Photometric Reproducibility: Certified Neutral Density Filters √ Stray Light Performance: Certified Stray Light Glass Cut Off Filters √ Spectral Bandwidth. √ Baseline Flatness. √ Drift.
British/European Pharmacopoeia Calibration - UV/Visible Spectrophotometers
√ Wavelength Verification: Certified Holmium Oxide Solution √ Photometric Accuracy: Certified Potassium Dichromate Solution √ Stray light performance: Certified Potassium Chloride Solution √ Resolution verification: Certified Toluene in Hexane Solution
Certified Reference Materials used to verify the performance of spectrophotometers are serialised and issued and with calibration certificates traceable to internationally recognised regulatory authorities.
Performance Verification - Fluorescence Spectrophotometers √ Sensitivity √ Raman Band of Water Measurements √ Drift & Noise
Standard Calibration - Absorption Microplate Readers
√ Verifying Wavelength Accuracy √ Verifying Optical Density √ Verifying Linearity & Reproducibility √ Checking Optical Alignment of Microplate Reader
Special test plates used to verify the performance of microplate readers are serialised and issued with data sheets traceable to internationally recognised regulatory authorities.
|
|
|